Surgical Kit

Beta and Gamma surgical kits are designed to contain and make immediately available, under sterile conditions, all the components for the surgical intervention.


  • One kit for all Beta and Gamma implants;
  • A simple tray, smart and practical;
  • Faster and well organized work thanks to an intuitive colour-code;
  • Wide range of placement of the surgical tray for an extra comfort;
  • The kit is modular and customizable according to the specific needs and preferences.
  • Cleaning

    01

    Remove all the pieces contained in the kit and dismantle those composed

    02

    Leave the pieces immersed for 15 minutes in a detergent solution

    03

    Wash the pieces with ultrasound for 15 minutes using an enzymatic detergent with neutral ph diluted in water (according to product instructions)

    04

    Rinse with water for 3 minutes

    Sterilization

    Dentists usually use sterilization with autoclave. Users must follow the instructions and the specific cycles recommended for the autoclave used.

    The operating cycles that are generally used are:

    • 121°C x 20 min. at 1 atm + drying
    • 134°C x 3,5 min. at 2 atm + drying

    Instructions for Use and Maintenance

    The kits are supplied in conditions of NON sterility, the user has therefore to arrange for the sterilization before use. Dry the kits with a soft cloth. Do not use abrasive sponges or metal parts which may damage the surface. Before sterilization is necessary to ensure that all the parts have been cleaned from the residual contaminants. The material may change its colour tone when exposed to UV rays for long periods. Keep the kit far from lights when not in use. The materials used can withstand a high number of sterilization cycles. it is recommended not to exceed the maximum values of tolerance estimated, which are: Maximum time: 20 min - maximum temperature 135 ° C and maximum pressure: 2.2 bar.